2020 STM 10
2011 -6330

#6330 Vaginal Electrode
$44.00

2001-6340

#6340 Rectal Electrode
$44.00

2011-5328

#5328 24"
#5338 6 ft
$45.00/$55.00

STM-10 Monthly Rental Fee $60.00 / $475.00 purchase price.
All rental fees go toward the purchase price.

Pathway STM-10 Pelvic Floor Electrical Muscle Stimulation System requires one of the above electrodes or cable set(s).

Federal Law (USA) restricts this device to sale by or on the order of a physician (or licensed practitioner).

PRODUCT FEATURES:

Programmable Settings:

  • SESSION TIME SETTINGS: (MIN) 5, 10, 15, 20, 25, 30
  • DUTY CYCLE SETTINGS: 5/5, 5/10, 10/10, 10/20 AND CONTINUOUS
  • STIM MODE SETTING: (Hz) 12.5(12), 50, 100, 200, 12:50 (12 followed by 50)
  • COMPLIANCE TIMER: hours: minutes (99:59 max)

Easy On/Off and Amplitude (intensity) keys for patient settings

Indications for Use:

  • Urinary Incontinence; Stress, Urge and Mixed Incontinence
  • Neuromuscular Reeducation

Contraindications:

  • Active infection or Genital Disease
  • Severe Pelvic Pain
  • Pregnancy
  • Postpartum or Post Surgical (6 weeks)
  • Atrophic Vaginitis
  • Dyspareunia
  • Menstrual Period
  • Pacemaker or Cardiac Arrhythmia
  • Presence of Any Known Malignancy
  • History of Severe Urine Retention
  • Prolapse
  • Diminished sensory perception

Treatment Protocols and Guidelines
Treatment Time:

Condition


Mild
Moderate
Severe

Weeks


1 - 8
1 - 10
1-16

Treatment Time


15/30 minutes
10/30 minutes
15/30minutes

Sessions/Day


1 or 2
1 or 2
1 or 2

The length of the treatment session is chosen to correspond with the above suggested treatment table, however, the session length should be adjusted accordingly using the following guidelines:

Guideline
To decrease muscle fatigue

Action
Decrease Session Length

AMPLITUDE:

Adjust to strong contractions but not uncomfortable. Allow a few minutes before final adjustment.

FREQUENCY:

The treatment frequency is contingent on the diagnosis. The following diagnosis corresponds to the following treatment frequencies:

Diagnosis


Stress Urinary Incontinence
Urge Urinary Incontinence

Frequency


50 Hz - 100 Hz
12.5 Hz

DUTY CYCLE:

The duty cycle is chosen based on the ability to allow the muscle to recover, which is determined by the clinician using the following guidelines:

Guideline


Strong Muscle Strength
Weak Muscle Strength
Induce Fatigue

Treatment Duty Cycle


1:1 Ratio
1:2 Ratio
(C) Continuous

Pathway STM-10 Technical Specifications:

  • Pulse Amplitude (mA): 1 - 99
  • Pulse Rate (Hz): 12.5/50/100/200
  • Cycle (on/off) (S): 5/5, 5/10, 10/10, 10/20, Continuous (C)
  • Ramp Up (S): 2 fixed
  • Ramp Down (S): 1 fixed
  • Pulse Width (uS): 300
  • Charge per pulse (uC): 17
  • Power Source: 9V alkaline battery
  • Dimensions (in): 4.75 X 2.5 X 1
  • Weight (oz.) 4.2 (5.8 w/ 9V alkaline battery)