#6330 Vaginal Electrode
$44.00

#6340 Rectal Electrode
$44.00

#5328 24"
#5338 6 ft
$45.00/$55.00
STM-10 Monthly Rental Fee $60.00 / $475.00 purchase price.
All rental fees go toward the purchase price.
Pathway STM-10 Pelvic Floor Electrical Muscle Stimulation System requires one of the above electrodes or cable set(s).
Federal Law (USA) restricts this device to sale by or on the order of a physician (or licensed practitioner).
PRODUCT FEATURES:
Programmable Settings:
- SESSION TIME SETTINGS: (MIN) 5, 10, 15, 20, 25, 30
- DUTY CYCLE SETTINGS: 5/5, 5/10, 10/10, 10/20 AND CONTINUOUS
- STIM MODE SETTING: (Hz) 12.5(12), 50, 100, 200, 12:50 (12 followed by 50)
- COMPLIANCE TIMER: hours: minutes (99:59 max)
Easy On/Off and Amplitude (intensity) keys for patient settings
Indications for Use:
- Urinary Incontinence; Stress, Urge and Mixed Incontinence
- Neuromuscular Reeducation
Contraindications:
- Active infection or Genital Disease
- Severe Pelvic Pain
- Pregnancy
- Postpartum or Post Surgical (6 weeks)
- Atrophic Vaginitis
- Dyspareunia
- Menstrual Period
- Pacemaker or Cardiac Arrhythmia
- Presence of Any Known Malignancy
- History of Severe Urine Retention
- Prolapse
- Diminished sensory perception
Treatment Protocols and Guidelines
Treatment Time:
Condition
Mild
Moderate
Severe
Weeks
1 - 8
1 - 10
1-16
Treatment Time
15/30 minutes
10/30 minutes
15/30minutes
Sessions/Day
1 or 2
1 or 2
1 or 2
The length of the treatment session is chosen to correspond with the above suggested treatment table, however, the session length should be adjusted accordingly using the following guidelines:
Guideline
To decrease muscle fatigue
Action
Decrease Session Length
AMPLITUDE:
Adjust to strong contractions but not uncomfortable. Allow a few minutes before final adjustment.
FREQUENCY:
The treatment frequency is contingent on the diagnosis. The following diagnosis corresponds to the following treatment frequencies:
Diagnosis
Stress Urinary Incontinence
Urge Urinary Incontinence
Frequency
50 Hz - 100 Hz
12.5 Hz
DUTY CYCLE:
The duty cycle is chosen based on the ability to allow the muscle to recover, which is determined by the clinician using the following guidelines:
Guideline
Strong Muscle Strength
Weak Muscle Strength
Induce Fatigue
Treatment Duty Cycle
1:1 Ratio
1:2 Ratio
(C) Continuous
Pathway STM-10 Technical Specifications:
- Pulse Amplitude (mA): 1 - 99
- Pulse Rate (Hz): 12.5/50/100/200
- Cycle (on/off) (S): 5/5, 5/10, 10/10, 10/20, Continuous (C)
- Ramp Up (S): 2 fixed
- Ramp Down (S): 1 fixed
- Pulse Width (uS): 300
- Charge per pulse (uC): 17
- Power Source: 9V alkaline battery
- Dimensions (in): 4.75 X 2.5 X 1
- Weight (oz.) 4.2 (5.8 w/ 9V alkaline battery)